Neeo S9 – S12


 

  • Neeo S available in different models: with stationary tube and Image Intensifier 9” or 12”  
  • Horizontal run 210 mm
  • Arc orbital Movement 130°
  • WIG-WAG ±12°
  • C-arm depth 659 mm
  • S. I. D. 730 mm
  • Arm rotation around horizontal axis ±270°
  • Generator power   3,5kW
  • Generator operating frequency 40 kHz
  • KV range 40 -h 110 kV
  • Max current in fluoroscopy 5 mA - 7 mA «pulsed» – 10 mA One Shot
  • Max current in «Boosted» 12 mA
  • Max current in radiography 65 mA
  • Maximum Current Absorbed: In fluoroscopy: 6 A – 230 Vac Fluoro Pulsed - In Radiography: 22 A – 230 Vac Radio Pulsed

Operating Modes and Functionality

OPERATING MODALITIES OF MEMORY

CONTINUOS FLUOROSCOPY
PULSED FLUOROSCOPY
ONE SHOT FLUOROSCOPY
CONTINUOS , PULSED AND ONE SHOT LOW DOSE FLUOROSCOPY
RADIOGRAPHY WITH CASSETTE
 

Image Itensifer and CCD

  • Useful Entrance Field Size: 215 mm / 160 mm / 120 mm
  • Contrast Ratio on large area: 23:1, 25:1, 30:1
  • Image resolution: 1000 x 1000 Pixels
  • Digital Video Output: 12 bits
  • Maximum frame speed: 30 fps
  • MTF: Motorized Lens ≥ 50%

Software:

Functions:Fluoroscopy 25 images per second 1024x 1024 x 12 bit; Recursive filter and movement detection; Acquisition rate 1,3,6,12, 25 image/sec; Electronic rotation with 1° step; Horizontal and vertical inversion; Brightness and contrast; Maximum opacity fluoroscopy acquisition; Real time subtraction with auto/manual mask; Storage on disk up to 25 images per second; Virtual collimators; Grey scale inversion; Spatial filters (edge enhancement); Cineloop; Landmarking; Overview 4-9-16 images; Print on Windows USB2 printers; Storage examination on removable drive (USB2 Pen drive) in BMP

Measure tools (included in basic configuration): - Distance Calculator - Angles - Stenosis - Overwrite text

  • 12" and 9" image intensifier option available
  • High Frequency monoblock X-ray generator 3.5 KW
  • 1k CCD camera delivers sharp and detail-rich images
  • Automatic” Technique (0 Points) with automatic control on kV and mA
  • Electronic zoom selection according to the number of fields of the Image Intensifer
  • Display touch screen (also working with work gloves) backlit and high contrast for the insertion of all parameters and for the display of any error messages or system issues
  • Full Dicom(optional)
  • foot-switch fluoroscopy control
  • 210 mm horizontal C-arm run, with manual brake for locking
  • Orbital movement 130°
  • 270° on each side arm rotation, with manual brake for locking.
  • 12° on each side C-arm swivelling with manual brake for locking. 
  • Monitor trolley with 2, 19 "high resolution monitors
  • Safety: - Filament current - mA min and mA max - Max Exposure Time - Max Temperature of the monobloc - Counting monoblock thermal units - Timer with RX stop every 5 minutes (up to 99 minutes) - max kV, min kV, max kV, max I - Anode Rotation - Microprocessor auto test
  • DICOM package:
    - DSA (Digital Subtraction Angiography) - QA (Quantitative Analysis) - HCF (High Contrast Fluoroscopy) - Dicom BASE Verify and Storage - Dicom PRINT Print on a DICOM Printer - Dicom WORKLIST - Dicom MPPS (Modality Performed Procedure Step) - Dicom STCOMMIT (Storage Commitment) - Dicom QUERY/RETRIEVE - Dicom CD/DVD
  • RTP Real Time Processing NW
  • Dosimetric system
  • Stainless steel cassette holder
  • Cross laser locator
  • Sony Thermal printer A4 format or A6 format
  • Sterile drapes